Transmille 1000-Series User Manual
Page 62
1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 62
9) Press and hold the Insulation tester TEST button and ensure it is kept depressed
for the duration of the test
10) Press
to activate the calibrator output. The LED next to the
key will light
up, as well as the terminal indicator; indicating that the output is active
11) The 1000 Series will show the insulation resistance value which can be read
from the tester (UUT) display
Note : If the polarity of the input from the insulation tester is reversed, the
1000 series will display INCORRECT INPUT. In this case, simply reverse
the input from the insulation tester to resolve this.