1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 62

9) Press and hold the Insulation tester TEST button and ensure it is kept depressed

for the duration of the test

10) Press

to activate the calibrator output. The LED next to the

key will light

up, as well as the terminal indicator; indicating that the output is active

11) The 1000 Series will show the insulation resistance value which can be read

from the tester (UUT) display

Note : If the polarity of the input from the insulation tester is reversed, the
1000 series will display INCORRECT INPUT. In this case, simply reverse

the input from the insulation tester to resolve this.