1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 48

Setting Passive Resistance Output

Note: The calibrator uses standard resistors of fixed decade values. The nearest

available resistance to the entered value will be automatically selected.


The 1000 series calibrator features passive resistance output as an alternative to
simulated resistance. This method does not allow the output to be varied, as the
values are fixed passive standards.

Complete the following example to set an output using the passive resistance
output.

1) Ensure the calibrator’s output has been set to Standby. This can be verified

by ensuring that standby is indicated on the display, and the output standby
status LED is lit.

2) Connect the UUT to the calibrator as described for resistance measurements

(see connection diagram on Page 30).

3) Select the correct range on the UUT.
4) Press

5) Press

6) Select the PASSIVE soft key.
7) The display will now show the 10 Ohm passive output (this is the lowest

available output).

8) To change the value, using the numeric keys enter the appropriate decade

value, e.g. 100

9) Press the multiplier key (if required) i.e.