Setting the Simulated Resistance Output
The calibrator defaults to simulated resistance output. This provides a fully variable
resistance output from 0 Ohms to 10M Ohms.
Complete the following example to set 10 Ohms output using the simulated
resistance output.
1) Ensure the calibrator’s output has been set to Standby. This can be verified
by ensuring that standby is indicated on the display, and the output standby
status LED is lit.
2) Connect the UUT to the calibrator as described for resistance measurements
(see page 30 for connection diagram)
3) Select the correct range on the UUT.
4) Press the numeric and decimal point keys to enter the required value, e.g.
1.234
5) Press the multiplier key (if required) e.g.
6) Press
7) The display will now indicate the value that has been entered below the
currently set output
8) Press
. The new value will replace the existing output in the middle
of the display