Insulation Test Voltage Measurement (Option)

The calibrator can measure insulation test voltage measurement from an
insulation/continuity or installation tester up to 1000V

1) Ensure the calibrator’s output has been set to Standby. This can be verified by

ensuring that standby is indicated on the display, and the output standby status
LED is lit.

2) Set the insulation tester to the required insulation voltage range and connect it to

the voltage terminals of the calibrator.
Note : For this function it is permitted to connect a voltage output device to the
calibrator terminals.

3) Press the INS TEST soft key. If the INS TEST key is not visible, press the

NEXT

key until the INS TEST key appears

4) Press V DOWN

or V UP to select the required voltage range as set on the UUT

5) Press and hold the Insulation tester TEST button and ensure it is kept depressed

for the duration of the test

6) Press

to activate the calibrator output. The LED next to the

key will light

up, as well as the terminal indicator; indicating that the output is active