Capacitance Commands

Function

Range

Command

Capacitance

10nF

R35

100nF

R38

1uF

R39

Standby Mode

Standby ON

S1

Standby OFF

S0


The capacitance section consists of a set of range commands which are used in
conjunction with the standby command. To enable a capacitance to be set, the
following command sequence should be used:

/

The use of the output (O) command is not necessary as the capacitance ranges are
fixed ranges at decade points.

For example, to set the 10nF output with the output switched on, the command is :

R35/S0

R35 = 10nF Range (as detailed in the table above)
S0 = Standby OFF (i.e. output switched ON)
= Carriage Return (ASCII character 13)