9) Press
to activate the calibrator output. The LED next to the
key will
light up, as well as the terminal indicator; indicating that the output is active.
To adjust the value, either type in a new value using the keyboard, for example 110
Ohms, or deviate the output using the cursor keys or the digital control.
Nulling the UUT
The value displayed/set on the calibrator is the value at the terminals. Therefore the
measuring instrument should be zeroed (Nulled) with the leads shorted before
connection to the calibrator.
This will remove any errors due to the resistance of the leads, especially at low
values where resistance of the test leads may be significant.