Setting an ac voltage output, Warning, Etting an – Transmille 1000-Series User Manual
Page 40: Oltage, Utput
1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 40
Setting an AC Voltage Output.
Complete the following procedure to set a AC voltage output. The
and
keys
can be used to edit the entry in the event of an incorrect key press.
WARNING
ENSURE THE OUTPUT DOES NOT EXCEED RATING FOR UUT INPUT
1) Ensure the calibrator’s output has been set to Standby. This can be verified
by ensuring that standby is indicated on the display, and the output standby
status LED is lit.
2) Connect the UUT to the calibrator as described for AC Voltage
measurements (see connection diagram on Page 29)
3) Select the correct range on the UUT.
4) Press
. The calibrator will switch to AC output.
5) Press the numeric and decimal point keys to enter the required voltage
output, e.g. 4.5678
6) Press the multiplier key (if required) e.g.
7) Press
8) The display will now indicate the value that has been entered below the
currently set output