1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 59

1) Ensure the calibrator’s output has been set to Standby. This can be verified

by ensuring that standby is indicated on the display, and the output standby

status LED is lit.

2) Connect the EA002 to the appropriate terminals

3) Ensure the coil platform is attached to the EA002 and that the clamp coil is

correctly aligned.

4) Press

5) Press

6) Press

7) Press the COIL soft key

8) The following menu will appear

9) Select the number of turns (e.g. 50)

10) Press SELECT or

11) Using the numerical and decimal keys, enter the required output (i.e. 500)

12) Press the

key, depending upon the polarity of the output required

(default is positive)

13) Press