Transmille 1000-Series User Manual
Page 37
1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 37
10) Press
. The new value will replace the existing output in the middle of
the display.
11) Press
to activate the calibrator output. The LED next to the
key will
light up, as well as the terminal indicator; indicating that the output is active.
The calibrator will now produce 56.789 mV DC at the voltage terminals. Once on a
range, any new output within that range can be set without the calibrator returning to
standby.
When a high voltage value is entered the calibrator will automatically go into standby
mode. To output the voltage, press the Output On key. This safety feature stops the
accidental output of high voltage.
For protection of UUT’s, the 1000 Series calibrators high voltage outputs are
ramped, so that voltage is increased gradually. Please refer to the ‘High Voltage
Output Ramp’ section (Page 69) for further information on this function.
For safety reasons the 1000 Series calibrator is fitted with a High Voltage timeout.
Please refer to the ‘High Voltage Timeout’ section (Page 69) for further information
on this function.
For safety reasons the 1000 Series calibrator is fitted with a High Voltage current
limit. This ensures that incorrect connections, faulty UUT’s or potentially dangerous
situations are protected against. Please refer to the ‘High Voltage Current Limit’
section (Page 70) for further information on this function.