Continuity Resistance Calibration (Option)
The 1000 Series provides a continuity resistance function for calibrating the
continuity ranges of insulation testers.
1) Ensure the calibrator’s output has been set to Standby. This can be verified by
ensuring that standby is indicated on the display, and the output standby status
LED is lit.
2) Set the tester to the required continuity range and connect it to the voltage
terminals of the calibrator.
3) Press the CONT. soft key. If the CONT. key is not visible, press the NEXT key
until the CONT. key appears
4) Continuity mode has two functions – the default function is resistance with a
current measurement function also available. To select between these two
functions press either the RESIST. soft key or the CURRENT softkey
5) To change the resistance value, using the numeric keys enter the required value,
e.g. 99.8
6) Press the multiplier key (if required) eg.
7) Press
8) Depending on the model of tester you may need to press the TEST button on the
tester to start measuring (this does not apply to all testers).
9) Press
to activate the calibrator output. The LED next to the
key will light
up, as well as the terminal indicator; indicating that the output is active