1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 87
For example, to set the 10 kHz output with the output switched on, the command
would be:
R58/H4/S0
R58 = Frequency Mode Activation (as detailed in the table above)
H4 = 10 kHz output (as detailed in the table above)
S0 = Standby OFF (i.e. output switched ON)
To set a frequency other than a preset the following command line would be used,
for example 15 kHz output with the output switched on:
R58/O15000/SO
R58 = Frequency Mode Activation (as detailed in the table above)
O15000 = 15 kHz output
S0 = Standby OFF (i.e. output switched ON)
Additional examples
1 Hz
R58/H0/S0
100 kHz R58/H7/S0
The calibrator will respond to the commands sent with the response codes as
detailed at the beginning of this section. These codes can be used to ensure that
hazardous output conditions are clearly indicated to the operator and to maintain
control of these outputs. This allows the calibrator to be returned to a safe state
once the testing required has been completed (e.g. Setting the calibrator back to
standby once a test is complete and ensuring this has been successfully achieved
and no hazardous outputs remain on the terminals).