2 Instrument Description
2.5.9
Response to instrument reconfiguration
When the instrument is reconfigured, the current counter/integrator recording
periods remain unaffected. The counter/integrator values will not be reset to 0
and the recording period will not be restarted.
2.5.10 Response to reconfiguration of the counters on the instrument
If a counter/integrator on the instrument is reconfigured, then only the affected
counter/integrator will be concluded, saved, reset to 0 and have the recording
period restarted.
If anything is changed through reconfiguration in the menu Configuration
Counters/integrators
General settings, then all counters/integrators are
concluded, saved, the values reset to 0, and the recording period is restarted.
2.5.11 Response to reconfiguration of counters with setup program
If a counter/integrator is reconfigured, the setup program will conclude and
save all counter states and reset the value to 0 and the recording period is
restarted.
H
The values can be deliberately reset, via the “Parameterization”
menu.