beautypg.com

Cleaning – West Control Solutions ProVU 4 User Manual

Page 17

background image

ProVU4 Controller, Profiler & Recorder/Controller - Product Manual

Page 16

Installation

59407, Issue 1 – March 2008

Instruments may be mounted side-by-side in a multiple installation, but instrument to panel
moisture and dust sealing will be compromised. Allow a 20mm gap above, below and behind
the instrument for ventilation. The cut-out width (for n instruments) is:

(96n - 4) mm or (3.78n - 0.16) inches

If panel sealing must be maintained, mount each instrument into an individual cut-out with
6mm or more clearance between the edges of the holes.

Note:

The mounting clamp tongues may engage the ratchets either on the sides or the
top/bottom faces of the Instrument housing. When installing several Instruments side-by-
side in one cut-out, use the ratchets on the top/bottom faces.

CAUTION:

Ensure the inside of the panel remains within the instrument operating temperature
and that there is adequate airflow to prevent overheating.

Gasket

Mounting Panel

Clamp

Ratchets

Instrument

Housing

Slide mounting clamp over

the instrument housing

towards rear face of

mounting panel until the

tongues engage in ratchets

and instrument is clamped in

position.

Hold instrument firmly in

position (apply pressure to

bezel only)

Figure 2. Panel-Mounting the instrument

CAUTION:

Do not remove the panel gasket, as this may result in inadequate clamping and
sealing of the instrument to the panel.

Once the instrument is installed in its mounting panel, it may be subsequently removed from
it’s housing, if necessary, as described in the Fitting and Removing Option Modules section.

Cleaning

Clean the front panel by washing with warm soapy water and dry immediately If the USB
option is fitted, close the USB port cover before cleaning.