Salinity Settings Page 2 – Hi-Lo Set Points

Figure 40: Salinity Settings Page 2 – Hi-Lo Set Points


This page allows you to set alarm limits for the salinity measurement mode:

Parameter

Description

Factory Default

Alarm Set

Points

Set alarm limits for the salinity measurement mode:
ENABLE – The measurement screen shows HI or LO
alarm indicators when the meter reading is above ‘HI
alarm value’ or below ‘LO alarm value’ specified in HI
& LO ppm parameters (below)
DISABLE – HI and LO alarms are not shown in the
LCD

DISABLE

Hi ppm

Specify Hi alarm value here. The Hi alarm occurs
when the salinity reading goes above this value

2.00 ppm

Lo ppt

Specify Lo alarm value here. The Lo alarm occurs
when the salinity reading goes below this value

(This parameter is not editable when ‘ALARM SET
POINT’ is disabled)

1.00 ppm

Press (Up) or (Down) arrow key to set high or low alarm point.

2. Salinity Calibration Mode

Use a standard solution of known value for calibration. Use the single-point
standardization option to measure samples that are close in value and multi-point
standardization to measure wide range of samples.

1. Switch the meter on. Make sure the meter is in salinity measurement mode.

2. Press CAL (F2) to go to calibration mode.

Note: If the meter is password protected, you will be prompted to enter a
password.