9 calibration report, 1 to view calibration report, Conductivity measurement mode – Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 79: Calibration report, To view calibration report

71
2.9 Calibration Report
Calibration report gives you detailed information on calibration. It includes date
& time, calibration points, range, factor and cell constant.
2.9.1 To View Calibration Report:
1. From conductivity measurement mode, press left or right arrow key to
navigate to other available functions until you see REPO function in the
LCD.
2. Press REPO (F2) key. The calibration report is shown in the display.
3. Press PRIN (F2) to transfer the calibration report to computer through IrDA.
Figure 33: Conductivity Calibration Report
Function Keys available in conductivity calibration report screen:
PRIN (F2)
Transfers calibration report to Computer through IrDA
NEXT (F3)
Goes to conductivity measurement mode
ESC (F4)
Goes to conductivity measurement mode
ENTER
(Not functional)
(Not functional)
3. Conductivity Measurement Mode
In conductivity measurement mode, the meter displays conductivity and temperature
readings. The LCD shows related information for the conductivity measurement such
as temperature compensation mode, stable indicator, probe condition, calibration
points, response time of the conductivity probe and conductivity alarm conditions.
You can customize or enable/disable some of these indicators in the Setup mode.
Note: Auto calibration can also be done for electrode with cell constant 0.4 – 2.