PART B – Concentration (mg/L) (ppm)
Mode

1. O2 mg/L (ppm) – DO Concentration Setup


DO Concentration

(mg/L) setup screen present many options to control the

operating parameters, which can be controlled and set from the DO setup
screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-
P (F1)
to navigate through these pages.

DO (mg/L) – Page 1

Figure 56 : DO (mg/L) – Page 1


This page allows you to set operating parameters for the DO Concentration
measurement mode:

Parameter

Function

Factory

Default

Select DO Unit

Sets the unit of measurement for DO

Available units: mg/L and ppm

mg/L

Auto Salinity Comp

DISABLE – No automatic compensate
ENABLE – The meter will apply measured salinity for DO
concentration of your sample.

Note: This option is activated only in CD650 & PCD650.

DISABLE

Set Salinity

Allows you to set the salinity (in ppt) for your solution

Available range: 0 to 50 ppt

0.0 ppt

NOTE: Auto salinity compensation will be applicable only in multi measurement
mode and only if one of the measurement selected is conductivity related. Otherwise,
a manually set salinity value will be applied.