1. pH Setup


pH setup sub-group allows you to configure pH measurement & calibration related
settings of the meter. The settings are displayed in 2 pages. Press NEXT-P (F2) and
PREV-P (F1) to navigate through these pages.

pH Settings Page 1 – pH Measurement & Calibration

Figure 15: pH Settings Page 1 – pH Measurement & Calibration


This page allows you to set pH measurement & calibration related settings of the
meter.

Parameter

Description

Factory Default

Buffer

Select your preferred pH standard buffer group for
calibration & auto-recognition.
Available groups: USA, NIST, DIN, PWB & USER
Select ‘USER’ if you need to use a custom buffer for
calibration.

USA

Cal points

Select the number of calibration points you intend to
calibrate the meter using the selected buffer group.
Available points: depends on the selected buffer group

3

Alarm Set

Point

ENABLE – The measurement screen shows HI or LO
alarm indicators when the meter reading is above ‘HI
alarm value’ or below ‘LO alarm value’ specified in HI
pH & LO pH parameters (below)
DISABLE – HI and LO alarms are not shown in the
LCD

DISABLE

Hi pH

Specify Hi alarm value here. The Hi alarm occurs
when the pH reading goes above this value
Available range: Hi pH (specified below) to 20.00
(This parameter is not editable when ‘ALARM SET
POINT’ is disabled)

20.00pH

Lo pH

Specify Lo alarm value here. The Lo alarm occurs
when the pH reading goes below this value
Available range: 0.00 to Lo pH (specified above)
(This parameter is not editable when ‘ALARM SET
POINT’ is disabled)

0.00pH