Thermo Fisher Scientific EC/TDS/SaltTestr User Manual
Ec/tds/salt testr, Instruction manual
I
NSTRUCTION MANUAL
EC/TDS/SALT Testr
Large Screen
Waterproof Conductivity/TDS/Salt Tester
Introduction
Thank you for selecting our microprocessor-based waterproof EC / TDS / SALT tester with large
dual line display. You have one of the following models:
• ECTestrlow
• ECTestrlow+
• ECTestrhigh
• ECTestrhigh+
• ECTestrpure+
• TDSTestrlow
• TDSTestrlow+
• TDSTestrhigh
• TDSTestrhigh+
• TDSTestrpure+
• SALTTestr
The non plus models are designed with user-replaceable two-pin type sensor with many user friendly
features such as the Hold function, Automatic Temperature Compensation and Self-Diagnostic Messaging
capabilities. The plus models, with the user-replaceable cup type sensor, have more additional
features such as simultaneous temperature display, °C or °F selection, and have higher resolution
measurement.
This manual provides a step-by-step guide to operate the testers.
Before you begin:
Remove the electrode’s protective cap. Soak the electrode for a few minutes in alcohol to remove any
oil stains on the electrodes which will affect the accuracy of the tester. Rinse thoroughly with de-
ionized water and shake off dry.
Calibration
The tester is factory calibrated. However, to ensure accuracy, you can calibrate the tester on a regular
basis. Depending on the model selected, you can calibrate the Conductivity / TDS / Salt / Temperature
on a one point calibration using the appropriate calibration standard.
The following table lists each model allowable calibration range when selecting a calibration standard.
It is best to select a standard close to the test solution value.
This Instruction Manual is also available for download on our Web-site: eutechinst.com or 4oakton.com
M o d e l
M o d e l
M o d e l
M o d e l
M o d e l
E C T
E C T
E C T
E C T
E C Te s t r
e s t r
e s t r
e s t r
e s t r l o
l o
l o
l o
l o wwwww
E C T
E C T
E C T
E C T
E C Te s t r
e s t r
e s t r
e s t r
e s t r l o
l o
l o
l o
l o w +
w +
w +
w +
w + E C T
E C T
E C T
E C T
E C Te s t r h i g h
e s t r h i g h
e s t r h i g h
e s t r h i g h
e s t r h i g h
E C T
E C T
E C T
E C T
E C Te s t r h i g h +
e s t r h i g h +
e s t r h i g h +
e s t r h i g h +
e s t r h i g h +
E C T
E C T
E C T
E C T
E C Te s t r
e s t r
e s t r
e s t r
e s t r p u r
p u r
p u r
p u r
p u r e +
e +
e +
e +
e +
Calibration
200 to 1990
200 to 1999
2.00 to 19.90
2.00 to 19.99
20.0 to 199.9
Standard Range
uS/cm
uS/cm
mS/cm
mS/cm
uS/cm
Model
Model
Model
Model
Model
TDST
TDST
TDST
TDST
TDSTestr
estr
estr
estr
estr lo
lo
lo
lo
lowwwww TDST
TDST
TDST
TDST
TDSTestr
estr
estr
estr
estr lo
lo
lo
lo
low+
w+
w+
w+
w+ TDST
TDST
TDST
TDST
TDSTestrhigh
estrhigh
estrhigh
estrhigh
estrhigh TDST
TDST
TDST
TDST
TDSTestrhigh+
estrhigh+
estrhigh+
estrhigh+
estrhigh+ TDST
TDST
TDST
TDST
TDSTestr
estr
estr
estr
estr pur
pur
pur
pur
pure+
e+
e+
e+
e+ S A L
S A L
S A L
S A L
S A LTTTTTestr
estr
estr
estr
estr
Calibration
200 to
200 to
1.00 to
1.00 to
20.0 to
1.00 to
Standard Range
1990 ppm
1999 ppm
10.00 ppt
10.00 ppt
199.9 ppm 10.00 ppt
Measurement
1. Press the ON/OFF button to switch the tester on.
2. Dip the electrode into the test solution making sure that it is fully immersed. Stir to clear any
trapped air bubbles from the electrode and let the reading stabilize. For plus models, you
can opt for the cup style measurement by filling the electrode cup with sample of test solution.
3. Note the value or press HOLD button to freeze the reading. To release the reading, press HOLD
again.
4. Press ON/OFF to turn off tester. If you do not press a button for 8.5 minutes, the tester will
automatically shut off to conserve batteries.
Figure 2: Example of a Power Up Sequence
Figure 3: Example of HOLD Function
Temperature Mode Selection (Only for PLUS Models)
This feature lets you select either the Celsius (°C) or the Fahrenheit (°F) temperature mode
display.
From the measurement mode,
1. Press °C/°F button to select the desired temperature mode. The temperature
display will toggle between the Celsius reading and the Fahrenheit reading.
Figure 4: Example of a temperature mode selection
1. Organise your calibration standard solution in two beakers – one for rinsing and
the other for calibration. Prepare separate de-ionized water for electrode rinsing.
2. Twist off and open the battery compartment lid (the cap with the lanyard loop).
You will see two switches – the INC (increment) and DEC (decrement) keys.
3. Switch on the unit using the ON/OFF key.
4. Rinse the electrode in de-ionized water and then rinse with the calibration
standard.
5. Dip the electrode to the calibration standard intended for calibration
and swirl gently to create a homogenous sample. Wait for the reading
to stabilize.
6. Press the INC or DEC key to enter calibration mode and to adjust the
reading to the calibration standard value. The upper display will be the
adjustable reading for calibration whereas the lower display will be the
default reading of the standard.
Note: At this point, the non-plus models will have its upper and lower display showing readings
with resolution of its plus version. Reading adjustment can then be made from its last significant
digit. Upon confirming the calibration, the reading changes back to its original resolution.
7. Wait for 5 seconds for the tester to automatically confirm the calibration value by displaying
“CO” and return to the measurement mode.
8. Replace the battery compartment lid. The tester is now ready for measurement.
Note: To exit this program without confirming the calibration, press the HOLD button before the
automatic confirmation takes place.
Note: The tester will exit calibration mode with “Er.0” message if the temperature is outside the
0 to 50°C range. Please refer to information under ’Self Diagnostic Messages’.
Note: The calibration window is +/- 50% from the default reading of the standard.
HOLD Function
This feature lets you freeze the display for a delayed observation.
1. Press HOLD button to freeze the measurement. A ‘HOLD’ indicator will be displayed and the
measurement will be frozen.
2. Press HOLD again to release the measurement. The ‘HOLD’ indicator will not be displayed
anymore indicating the held measurement is released.
Figure 1: Example of a calibration sequence using 2764 µS calibration standard
Table 1
+
-
INC Key
TDS Factor Setting (Only for TDS Models)
The TDS models features includes the setting of the TDS factor. You can adjust the TDS factor from 0.4
to 1.0.
From the measurement mode,
1. Press HOLD button to bring the tester to the HOLD mode.
2. Press either the INC or DEC key to enter the TDS Factor Setting mode. The upper display will
be the adjustable TDS factor whereas the lower display will be the last TDS Factor setting.
3. Use the INC or DEC key to set the correct TDS factor for your sample measurement.
4. Wait for 5 seconds for the tester to automatically confirm the new setting by displaying “CO”
and return to the measurement mode.
Figure 5: Example of setting the TDS Factor
Electrode Maintenance
1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized water and
wipe them dry with clean cloth before storing with its protective cap. For the plus models,
remove the white plastic cup insert to thoroughly clean viscous solutions. Never scratch
the electrodes with a hard substance.
2. For better performance, soak the electrode in alcohol for 10 to 15 minutes and rinse with
de-ionized water before starting any measurement process. This is to remove dirt and oil
stains on the electrode which may affect the accuracy of the measurements.
After 5 sec,
“CO” is
displayed
signifying
automatic
confirmation
INC
DEC
INC
DEC
After 5 sec,
“CO” is
displayed
signifying
automatic
confirmation
INC
DEC
INC
DEC
ON/OFF