Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 38

30
Resistivity
Range
Measuring ranges
Resolution
Sub range Accuracy
5
4
3
2
1
10.00MΩ to 20.00MΩ
1.000MΩ to 9.999MΩ
500.0KΩ to 999.9KΩ
100.0KΩ to 500.0KΩ
10.00KΩ to 99.99KΩ
3.333KΩ to 9.999KΩ
1.000KΩ to 3.333KΩ
250.0Ω to 999.9Ω
100.0Ω to 250.0Ω
25.00Ω to 99.99Ω
10.00Ω to 25.00Ω
2.000Ω to 9.990Ω
0.01MΩ
0.001MΩ
0.1KΩ
0.1Ω
0.01Ω
0.001Ω
0.001Ω
0.1Ω
0.1Ω
0.01Ω
0.01Ω
0.01Ω*
* Display resolution is 0.001 but actual
resolution shows 0.01.
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
1% of FS
Conductivity / TDS / Salinity /
Resistivity
COND 600
COND 610
Conductivity Range
0 to 200mS
0 to 500 mS
Salinity
-
80 ppt
Resistivity
-
0 to 20.00MΩ
TDS
200 ppt (depending on TDS factor)
500 ppt (depending on TDS factor)
Conductivity Cell constant
0.010 to 10.000
Conductivity Cell
2 & 4 Cell
TDS Conversion Factor
0.400 to 1.000
Temperature Comp.
Linear
Linear & Pure
Cal-Auto/Manual
Yes
Cal-Single/Multi
Yes
Cal Due Alarm
Yes (max-30 days)
Set Point Alarm
Yes
Input Conductivity
8 Pin Round
Dissolved Oxygen
DO 600, PD 650, CD 650, PCD 650
DO Range
0 to 600 %/ 90 mg/l
Resolution
0.1%, 0.01 mg/L
Accuracy
± 2%, ± 0.2 mg/L
Dissolved Oxygen Probe
Galvanic
Mode
% Sat, mg/L, ppm
Temperature Comp.
Linear
Barometric Pressure Compensation
Automatic
Barometric Pressure Range
450 to 825 mmHg
Barometric Pressure Resolution
1 mmHg
Barometric Pressure Accuracy
± 1%
Salinity Correction
0 to 50 ppt
Cal Due Alarm
Yes (max -30 days)