O2 % - do saturation setup, Part a - % saturation mode – Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 107
![background image](/manuals/370630/107/background.png)
99
PART A - % Saturation Mode
1. O2 % - DO Saturation Setup
The DO saturation (%) setup screen presents many options to control the
operating parameters, which can be controlled and set from the DO setup
screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-
P (F1) to navigate through these pages.
DO (%) Settings Page 1 – Operating Parameters
Figure 49 : DO (%) – Page 1
Parameter
Function
Factory
Default
Offset Cal
Allows to adjust % saturation offset calibration
0.0 %
Calibration due
Specify number of days for the DO calibration alarm. The
meter shows CAL DUE indicator after calibration date.
Available range : 0 to 30
5 days
Select Pressure Unit Sets the unit for barometric pressure
Available units: mmHg and kPa
mmHg
Measured Pressure Barometric pressure as measured by the instrument.
Adjusted Pressure User adjusted value.
-
Pressure
Compensation
ENABLE – The meter will compensate for the barometric
pressure at the location depending on the altitude.
DISABLE – No compensation for the pressure.
(This is applicable only for the % saturation mode. For mg/L
or ppm mode pressure comp. would always be applicable)
ENABLE
NOTE: During measurement, the meter will automatically measure and compensate
for barometric pressure. However, if the user feels that barometric pressure is
inaccurate and needs to be calibrated, it can be adjusted in the DO setup screen.