O2 % - do saturation setup, Part a - % saturation mode – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 107

background image

99

PART A - % Saturation Mode

1. O2 % - DO Saturation Setup


The DO saturation (%) setup screen presents many options to control the
operating parameters, which can be controlled and set from the DO setup
screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-
P (F1)
to navigate through these pages.

DO (%) Settings Page 1 – Operating Parameters

Figure 49 : DO (%) – Page 1

Parameter

Function

Factory

Default

Offset Cal

Allows to adjust % saturation offset calibration

0.0 %

Calibration due

Specify number of days for the DO calibration alarm. The
meter shows CAL DUE indicator after calibration date.
Available range : 0 to 30

5 days

Select Pressure Unit Sets the unit for barometric pressure

Available units: mmHg and kPa

mmHg

Measured Pressure Barometric pressure as measured by the instrument.

Adjusted Pressure User adjusted value.

-

Pressure

Compensation

ENABLE – The meter will compensate for the barometric
pressure at the location depending on the altitude.
DISABLE – No compensation for the pressure.
(This is applicable only for the % saturation mode. For mg/L
or ppm mode pressure comp. would always be applicable)

ENABLE

NOTE: During measurement, the meter will automatically measure and compensate
for barometric pressure. However, if the user feels that barometric pressure is
inaccurate and needs to be calibrated, it can be adjusted in the DO setup screen.