1 concentration calibration report, 1 to view calibration report, 2 set salinity – Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 118: Concentration calibration report, To view calibration report, Set salinity
![background image](/manuals/370630/118/background.png)
110
7. Press ENTER key to confirm the entered value.
8. Press ESC (F4) to go to measurement mode. The meter shows the
concentration reading of the solution in measurement mode.
Figure 59 : Concentration Calibration
2.1 Concentration Calibration Report
Calibration report gives you information on calibrated temperature and
calibration at pressure and salinity. It includes date & time on which the
last calibration was done and the offset.
2.1.1 To View Calibration Report:
1. From concentration measurement mode, press left or right
arrow key to navigate to other available functions until you see
REPO function in the LCD.
2. Press REPO (F2) key. The calibration report is shown in the display
Press PRIN (F2) to transfer the calibration report to the computer.
Figure 60 : DO Calibration Report in Concentration Mode
2.2 Set Salinity
This option allows you to set the salinity (in ppt) for your solution, as the
dissolved salts alter the relationship between the partial pressure of
oxygen and the oxygen concentration in water. By setting the salinity,
the meter will compensate for the impact of salinity on the partial
Note: The function of the keys available on the concentration calibration/
report screen is same as previously described in saturation calibration.