1 concentration calibration report, 1 to view calibration report, 2 set salinity – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 118: Concentration calibration report, To view calibration report, Set salinity

background image

110

7. Press ENTER key to confirm the entered value.

8. Press ESC (F4) to go to measurement mode. The meter shows the

concentration reading of the solution in measurement mode.

Figure 59 : Concentration Calibration

2.1 Concentration Calibration Report

Calibration report gives you information on calibrated temperature and
calibration at pressure and salinity. It includes date & time on which the
last calibration was done and the offset.

2.1.1 To View Calibration Report:

1. From concentration measurement mode, press left or right

arrow key to navigate to other available functions until you see
REPO function in the LCD.

2. Press REPO (F2) key. The calibration report is shown in the display

Press PRIN (F2) to transfer the calibration report to the computer.






Figure 60 : DO Calibration Report in Concentration Mode

2.2 Set Salinity

This option allows you to set the salinity (in ppt) for your solution, as the
dissolved salts alter the relationship between the partial pressure of
oxygen and the oxygen concentration in water. By setting the salinity,
the meter will compensate for the impact of salinity on the partial

Note: The function of the keys available on the concentration calibration/
report screen is same as previously described in saturation calibration.