2.3 Temperature Calibration for MTC mode

Make sure you have selected ‘MTC’ and required unit of measurement (ºC or
ºF) in Temperature settings. Switch on the meter. Make sure the meter is in
measurement mode.

1. Press CAL (F2) to go to calibration mode.

2. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds

and then shows the cell constant adjustment screen.

3. Press NEXT (F3) key. (For conductivity mode only. In pH mode, pressing

this key will take you to the Report page)

4. Press TEMP (F1) to go to temperature calibration. The temperature

calibration screen appears.

5. The screen shows two readings. The upper display shows the temperature

reading of the solution with respect to previous calibration (if any) & lower
displays shows the temperature reading of the solution without any
calibration (default reading). Use & keys to adjust the upper display to
the temperature reading of the thermometer.

6. Press ENTER key to confirm temperature value.

Function Keys available in temperature calibration screen:

NEXT (F3)

Goes to measurement mode from where you entered calibration

ESC (F4)

Goes to measurement mode from where you entered calibration

ENTER

Confirms calibration

Increase/decrease temperature reading

(Not functional)

3. Temperature Measurement

There is no page in the meter dedicated to temperature measurement mode as the
temperature is displayed with all other main parameters.


- END OF TEMPERATURE SECTION -

Note: The meter allows you to adjust the upper display reading to any
value within the measuring range

-10.0 ºC to 110.0 ºC (14.0 ºF to 230.0 ºF).