3 temperature calibration for mtc mode, Temperature measurement, Temperature calibration for mtc mode – Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 47

39
2.3 Temperature Calibration for MTC mode
Make sure you have selected ‘MTC’ and required unit of measurement (ºC or
ºF) in Temperature settings. Switch on the meter. Make sure the meter is in
measurement mode.
1. Press CAL (F2) to go to calibration mode.
2. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds
and then shows the cell constant adjustment screen.
3. Press NEXT (F3) key. (For conductivity mode only. In pH mode, pressing
this key will take you to the Report page)
4. Press TEMP (F1) to go to temperature calibration. The temperature
calibration screen appears.
5. The screen shows two readings. The upper display shows the temperature
reading of the solution with respect to previous calibration (if any) & lower
displays shows the temperature reading of the solution without any
calibration (default reading). Use & keys to adjust the upper display to
the temperature reading of the thermometer.
6. Press ENTER key to confirm temperature value.
Function Keys available in temperature calibration screen:
NEXT (F3)
Goes to measurement mode from where you entered calibration
ESC (F4)
Goes to measurement mode from where you entered calibration
ENTER
Confirms calibration
Increase/decrease temperature reading
(Not functional)
3. Temperature Measurement
There is no page in the meter dedicated to temperature measurement mode as the
temperature is displayed with all other main parameters.
- END OF TEMPERATURE SECTION -
Note: The meter allows you to adjust the upper display reading to any
value within the measuring range
-10.0 ºC to 110.0 ºC (14.0 ºF to 230.0 ºF).