Salinity setup – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 91

background image

83

1. Salinity Setup


Salinity setup screen presents options to control the operating parameters, which can
be controlled and set from the salinity setup screen. The settings are displayed in 2
pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.

Salinity Settings Page 1 – Operating Parameters














Figure 39: Salinity Settings - Page 1



Parameter

Description

Factory

Default

Pure water

Coeff.

Pure water coefficient will be calculated and applied automatically
for ultra pure water measurement if enabled.

N/A

(fixed)

Lnr Temp

Coefficient

Specify temperature coefficient to measure temperature
corrected conductivity.

N/A

(fixed)

Normalization

Temp

Select a default temperature that the meter will use along with the
temperature coefficient to calculate a temperature corrected
conductivity measurement.

15

(fixed)

Cell Constant

Select the appropriate cell constant of the conductivity cell.
Available cell constants: 0.1 to 10

1.000

Calibration

due

Specify number of days for the Conductivity calibration alarm.
The meter shows CAL DUE indicator after calibration date.
Available range: 0 to 30 (if enabled)

DISABLE

Calibration

points

Single - In single point calibration one calibrated conductivity
value can be used for the entire 5 conductivity ranges.
Multi - In multi point calibration, you can calibrate one point in
each of the measuring ranges (up to 5 points).

Single