1. Salinity Setup
Salinity setup screen presents options to control the operating parameters, which can
be controlled and set from the salinity setup screen. The settings are displayed in 2
pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
Salinity Settings Page 1 – Operating Parameters
Figure 39: Salinity Settings - Page 1
Parameter
Description
Factory
Default
Pure water
Coeff.
Pure water coefficient will be calculated and applied automatically
for ultra pure water measurement if enabled.
N/A
(fixed)
Lnr Temp
Coefficient
Specify temperature coefficient to measure temperature
corrected conductivity.
N/A
(fixed)
Normalization
Temp
Select a default temperature that the meter will use along with the
temperature coefficient to calculate a temperature corrected
conductivity measurement.
15
(fixed)
Cell Constant
Select the appropriate cell constant of the conductivity cell.
Available cell constants: 0.1 to 10
1.000
Calibration
due
Specify number of days for the Conductivity calibration alarm.
The meter shows CAL DUE indicator after calibration date.
Available range: 0 to 30 (if enabled)
DISABLE
Calibration
points
Single - In single point calibration one calibrated conductivity
value can be used for the entire 5 conductivity ranges.
Multi - In multi point calibration, you can calibrate one point in
each of the measuring ranges (up to 5 points).
Single