beautypg.com

Resistivity calibration mode – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 101

background image

93

2. Resistivity Calibration Mode

Use a standard solution of known value for calibration. Use the single-point

standardization option to measure samples that are close in value and multi-point
standardization to measure wide range of samples.

1. Switch the meter on. Make sure the meter is in resistivity measurement

mode.

2. Press CAL (F2) to go to calibration mode.

3. The meter shows the ‘Calibration-Rinse Electrode’ screen for a few seconds

and then shows the cell constant adjustment screen.

Figure 46 : Resistivity Cell Constant & Resistivity Calibration

4. The upper display of the resistivity cell constant screen shows the cell

constant of your cell & lower display shows the measured value of the
solution with adjusted cell constant. Press &

keys to adjust the cell

constant for your custom solution in upper display.

5. Press ENTER key to confirm the entered value and to go to resistivity

calibration screen.

6. The screen shows two readings. The upper display shows the resistivity

reading of the solution with respect to previous calibration (if any) & lower
display shows the resistivity reading of the solution without any calibration.
Use

&

keys to adjust the upper display of the resistivity value of your

custom solution.

7. Press ENTER key to confirm the entered value.

Note: If the meter is password protected, you will be prompted to enter a
password.