Ion setup, Ion calibration mode, 1 about ion calibration – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 67: Ion setup 2. ion calibration mode, About ion calibration

background image

59

1. Ion Setup


Ion setup sub-group allows you to configure Ion measurement & calibration related
settings of the meter.

Ion Settings Page











Figure 23: Ion Settings Page

Parameter

Description

Factory Default

Ion unit

Sets the unit of measurement for Ion.
Available units: ppm, molar and mg/L

ppm

Press (Up) or (Down) arrow key to change the value of a selected parameter.


2. Ion Calibration Mode

2.1 About Ion Calibration

The meter supports Ion calibration up to 8 points with minimum of 2-points. The
available 8-points are 0.001, 0.01, 0.1, 1, 10, 100, 1000 & 10000.

The meter guides your through the calibration process with on-screen
instructions and hints. If the second point and subsequent points are not within
the allowable calibration window (15mV/decade to 90mV/decade), the meter will
not accept the calibration.

When you try to re-calibrate the meter with ISE, the meter gives you 2 options:

Clear the previous calibration (choose this if you wish to delete all

previously calibrated points and re-calibrate with a new type of Ion)


Retain the previous calibration (choose this if you wish to retain

previously calibrated points and re-calibrate new points or overwrite
existing points with same type of Ion).