5.2.1 Logging data automatically in the meter’s memory

1. Make sure that the Print mode is set to ‘data logging mode’ in the System

Setup. You can also specify the time interval for each transfer.

2.

From measurement mode, press PRIN (F3) to start saving data
automatically in meter’s memory. The memory location of the transferred
reading is shown in the bottom-left of the screen.












5.2.2

Storing a current measurement

(In IrDA and LED print mode: Applicable to all

modes)

1. Make sure you are in the measurement mode.

2. Press left or right arrow key to navigate to other available functions

until you see STOR function in the LCD.

3. Press STOR (F3) key to store the current reading. The bottom left of the

screen will show the memory location where this reading is being saved.

5.2.3 Viewing stored data

1. Make sure you are in measurement mode.

2. Press left or right arrow key to navigate to

other available functions until you see MEM
function in the LCD.

3. Press MEM (F2) key to view the stored data.

The last data entry stored is shown in the
display with its memory location at the top-right
corner of the screen [Immediately under the
time]. (See following diagram)

4. To navigate to a particular memory location,

press the up or down arrow key to select it
and then press the ENTER key. The meter will
show the stored data in the memory location you selected.

5. Press NEXT (F3) key to return to the measurement mode.

6. Press ESC (F4) key to return to the main screen of the measurement mode.

Memory Location
where readings are
being saved

.

NOTE:

This symbol will

appear if data logging mode
has been selected in “System
set up’. It will automatically
start saving data in meter’s
memory, when PRIN (F3) key