beautypg.com

4 membrane/o-ring replacement (optional procedure), Membrane/o-ring replacement (optional procedure) – Thermo Fisher Scientific CyberScan DO 600 User Manual

Page 123

background image

115

Figure 63:

Positioning of O-

ring & Membrane

Close

Open

Installation tool

Insert installation
tool into slots on
membrane lock.
Then unscrew
membrane lock
from membrane
cap

Figure 64: Use tool to take

out (or put in) membrane

3.2.4 Membrane/O-ring Replacement (Optional Procedure)

It is recommended only experienced service personnel can
perform this procedure.

This procedure is OPTIONAL, and should only be performed if you
have new membrane and O-ring. You are also required to have a
membrane installation tool. These items are available as optional
accessories in the “Accessories” section.

1. Pull off the probe guard.

2. Unscrew the membrane cap from the probe.

3. Hold the probe under hot running water and brush away the

white oxide on the cylindrical anode with a stiff plastic brush –
do not use metal cleaning material.

4. If the cathode has any deposits, remove them with a light

scouring powder. Do not polish the cathode.

5. Using the installation tool, unscrew and remove the membrane

lock from the membrane cap. See Figure 64 on the following
page.

6. Remove the membrane and O-ring. Discard both.

7. Rinse the membrane cap and membrane lock in tap water.

8. Install a new O-ring inside the membrane cap.

9. Install a new membrane. Make sure the membrane covers

the O-ring all around its circumference.

10. Using the installation tool, screw the membrane lock back into the

cap. Tighten the lock firmly over the membrane and O-ring, but do
not over tighten.

11. Inspect the membrane for wrinkles. If wrinkles exist, remove the

membrane and repeat steps 8 – 11.

12. Fill the membrane cap with water and inspect the bottom for leaks.

If water drops are leaking from the membrane, re-seal
the membrane on the O-ring (repeat steps 8 – 11, for
membrane replacement only).

13. If the assembly is leak-free, empty the water and fill the

membrane cap with electrolyte to the brim.

14. Tap the side of the housing gently to remove any air

bubble that may be sticking to the membrane.

15. Screw the cap onto the probe. Excess electrolyte will

drain out.

16. Replace probe guard.

17. Calibrate the probe after the % saturation readings

have stabilised.