KROHNE MFC 010 C Converter User Manual
Page 88

86
MFC010 Interface Manual
Bit
Warning Flag Designation
8
Name
:
Zero Flow Calibration Above Limits
Cause
:
The Mass flow rate during the previous Mass flow zero calibration
attempt was above 10% of the nominal flow or this sensor.
9
Name
:
Zero Flow Calibration Failed
Cause
:
The previous Mass flow zero calibration failed. Refer to section
8.1 on page 71 for further details.
10
CT
Name
:
Temperature Drift
Cause
:
The measured temperature has exceeded ±30°C from the
temperature reference taken during the last Mass Flow zero
calibration. For use in Custody transfer applications.
11
CT
Name
:
Power Fail
Cause
:
This flag is set when the system is started to indicate that the
power supply has been interrupted.
12
CT
Name
:
Sampling
Cause
:
The tube oscillation control system has had difficulty locking onto
the tube drive frequency, which may have resulted in inaccurate
measurement values.
13
Name
:
2 Phase Signal Above Limit
Cause
:
2 phase signal is above the threshold limit set
14
Name
:
Tube Asymmetry
Cause
:
Not Currently Implemented
15
Name
:
External Vibration
Cause
:
Not Currently Implemented
16
Name
:
Pulsating Flow
Cause
:
Not Currently Implemented
17
Name
:
Partially Empty System
Cause
:
Not Currently Implemented
18
Name
:
Reserved for Future Use
19
Name
:
Reserved for Future Use
20
Name
:
Reserved for Future Use
21
Name
:
Reserved for Future Use
22
Name
:
Reserved for Future Use
23
Name
:
Reserved for Future Use
24
Name
:
Reserved for Future Use
25
Name
:
Reserved for Future Use