MFC010 Interface Manual

Bit

Warning Flag Designation

8

Name

:

Zero Flow Calibration Above Limits

Cause

:

The Mass flow rate during the previous Mass flow zero calibration

attempt was above 10% of the nominal flow or this sensor.

9

Name

:

Zero Flow Calibration Failed

Cause

:

The previous Mass flow zero calibration failed. Refer to section

8.1 on page 71 for further details.

10

CT

Name

:

Temperature Drift

Cause

:

The measured temperature has exceeded ±30°C from the

temperature reference taken during the last Mass Flow zero

calibration. For use in Custody transfer applications.

11

CT

Name

:

Power Fail

Cause

:

This flag is set when the system is started to indicate that the

power supply has been interrupted.

12

CT

Name

:

Sampling

Cause

:

The tube oscillation control system has had difficulty locking onto

the tube drive frequency, which may have resulted in inaccurate

measurement values.

13

Name

:

2 Phase Signal Above Limit

Cause

:

2 phase signal is above the threshold limit set

14

Name

:

Tube Asymmetry

Cause

:

Not Currently Implemented

15

Name

:

External Vibration

Cause

:

Not Currently Implemented

16

Name

:

Pulsating Flow

Cause

:

Not Currently Implemented

17

Name

:

Partially Empty System

Cause

:

Not Currently Implemented

18

Name

:

Reserved for Future Use

19

Name

:

Reserved for Future Use

20

Name

:

Reserved for Future Use

21

Name

:

Reserved for Future Use

22

Name

:

Reserved for Future Use

23

Name

:

Reserved for Future Use

24

Name

:

Reserved for Future Use

25

Name

:

Reserved for Future Use