MFC010 Interface Manual

Register No.

Description

1013

CT

Sensor Size

- The interpretation of this register depends upon the setting of the Sensor

Type register (Register Number 1012). WARNING: Changing this value will

reset the calibration coefficients.

Format : Unsigned Integer
Range :

For the OPTIMASS 7000 0 = Size 06

1 = Size 10

2 = Size 15

3 = Size 25

4 = Size 40

5 = Size 50

6 = Size 80

For the OPTIMASS 3000 0 = Size 01

1 = Size 03

2 = Size 04

For the OPTIGAS 5000

0 = Size 15

1 = Size 25

For the OPTIMASS 8000 0 = Size 15

1 = Size 25

2 = Size 40

3 = Size 80

4 = Size 100

For the OPTIMASS 9000 0 = Size 15

1 = Size 25

2 = Size 40

3 = Size 80

4 = Size 100

For the OPTIMASS 1000 0 = Size 15

1 = Size 25

2 = Size 40

3 = Size 50

For the OPTIMASS 2000 0 = Size 100

1 = Size 150

2 = Size 250

For the OPTIGAS 4000

0 = Size 15

Default Value : Dependent on meter supplied

1014

CT

Sensor Material

- The material from which the measuring tube is fabricated determines the

oscillation frequency and influences the calibration coefficient values. This

setting is limited by the sensor type setting (see Register 1012) as not all

material t ypes m ay b e av ailable i n al l s ensor f amilies. WARNING:

Changing this value will reset the calibration coefficients.

Format : Unsigned Integer
Range :

0 = Titanium (Available on OPTIMASS 7000 & 8000 Only)

1 = Hastelloy (Available on Optimass 7000, 3000, 8000 & 9000 Only)

2 = Stainless Steel

3 = Tantalum (Available on OPTIMASS 7000 Only)
Default Value : Dependent on meter supplied