KROHNE MFC 010 C Converter User Manual
Page 54
52
MFC010 Interface Manual
Register No.
Description
1013
CT
Sensor Size
- The interpretation of this register depends upon the setting of the Sensor
Type register (Register Number 1012). WARNING: Changing this value will
reset the calibration coefficients.
Format : Unsigned Integer
Range :
For the OPTIMASS 7000 0 = Size 06
1 = Size 10
2 = Size 15
3 = Size 25
4 = Size 40
5 = Size 50
6 = Size 80
For the OPTIMASS 3000 0 = Size 01
1 = Size 03
2 = Size 04
For the OPTIGAS 5000
0 = Size 15
1 = Size 25
For the OPTIMASS 8000 0 = Size 15
1 = Size 25
2 = Size 40
3 = Size 80
4 = Size 100
For the OPTIMASS 9000 0 = Size 15
1 = Size 25
2 = Size 40
3 = Size 80
4 = Size 100
For the OPTIMASS 1000 0 = Size 15
1 = Size 25
2 = Size 40
3 = Size 50
For the OPTIMASS 2000 0 = Size 100
1 = Size 150
2 = Size 250
For the OPTIGAS 4000
0 = Size 15
Default Value : Dependent on meter supplied
1014
CT
Sensor Material
- The material from which the measuring tube is fabricated determines the
oscillation frequency and influences the calibration coefficient values. This
setting is limited by the sensor type setting (see Register 1012) as not all
material t ypes m ay b e av ailable i n al l s ensor f amilies. WARNING:
Changing this value will reset the calibration coefficients.
Format : Unsigned Integer
Range :
0 = Titanium (Available on OPTIMASS 7000 & 8000 Only)
1 = Hastelloy (Available on Optimass 7000, 3000, 8000 & 9000 Only)
2 = Stainless Steel
3 = Tantalum (Available on OPTIMASS 7000 Only)
Default Value : Dependent on meter supplied