KROHNE MFC 010 C Converter User Manual
Page 69
MFC010 Interface Manual
67
Register No.
Description
3097 / 3098
Sensor Maximum Temperature Specification
- The m aximum t emperature r ating f or t he S ensor. The value is scaled
according to the setting of the “Temperature Units” (see Holding Register
No.1025)
Format : Floating Point
Range : 0 °C to + 500 °C
Default Value : Dependant on Sensor Settings (See Appendix D)
3099 / 3100
Sensor Minimum Temperature Specification
- The minimum temperature rating for the Sensor. The value is scaled
according to the setting of the “Temperature Units” (see Holding Register
No.1025)
Format : Floating Point
Range : -200 °C to + 50 °C
Default Value : Dependant on Sensor Settings (See Appendix D)
3101 / 3102
CT
Pressure Suppression Time
- The period of application of the Pressure Suppression, during which the
Pressure Suppression Cut-off is applied (see holding register 3103). S ee
section 8.10 on page 79 for further details of Pressure Suppression.
Format : Floating Point
Range : 0.0 to 20.0 Seconds
Default Value : 0.0 Seconds
3103 / 3104
CT
Pressure Suppression Cut-off
- When Pressure Suppression is active, the Cut-off value that is applied in
addition to the Low Flow Threshold (see Holding Register No. 3047). Se e
section 8.10 on page 79 for further details of Pressure Suppression.
Format : Floating Point
Range : 0.0% to 10.0%
Default Value : 0.0 %
3105 / 3106
Density Averaging
- The filter period of the Density reading.
Format : Floating Point
Range : 1.0 to 20.0 Seconds
Default Value : 10.0 Seconds
3107 / 3108
Concentration 2 Offset
- A fixed offset that the user can apply any of the Concentration 2
calculations made by the MFC010. S ee Section 8.4 on page 76 for more
details.
Format : Floating Point
Range : -10% to +10%
Default Value : 0 %