Register No.

Description

3097 / 3098

Sensor Maximum Temperature Specification

- The m aximum t emperature r ating f or t he S ensor. The value is scaled

according to the setting of the “Temperature Units” (see Holding Register

No.1025)

Format : Floating Point
Range : 0 °C to + 500 °C
Default Value : Dependant on Sensor Settings (See Appendix D)

3099 / 3100

Sensor Minimum Temperature Specification

- The minimum temperature rating for the Sensor. The value is scaled

according to the setting of the “Temperature Units” (see Holding Register

No.1025)

Format : Floating Point
Range : -200 °C to + 50 °C
Default Value : Dependant on Sensor Settings (See Appendix D)

3101 / 3102

CT

Pressure Suppression Time

- The period of application of the Pressure Suppression, during which the

Pressure Suppression Cut-off is applied (see holding register 3103). S ee

section 8.10 on page 79 for further details of Pressure Suppression.

Format : Floating Point
Range : 0.0 to 20.0 Seconds
Default Value : 0.0 Seconds

3103 / 3104

CT

Pressure Suppression Cut-off

- When Pressure Suppression is active, the Cut-off value that is applied in

addition to the Low Flow Threshold (see Holding Register No. 3047). Se e

section 8.10 on page 79 for further details of Pressure Suppression.

Format : Floating Point
Range : 0.0% to 10.0%
Default Value : 0.0 %

3105 / 3106

Density Averaging

- The filter period of the Density reading.

Format : Floating Point
Range : 1.0 to 20.0 Seconds
Default Value : 10.0 Seconds

3107 / 3108

Concentration 2 Offset

- A fixed offset that the user can apply any of the Concentration 2

calculations made by the MFC010. S ee Section 8.4 on page 76 for more

details.

Format : Floating Point
Range : -10% to +10%

Default Value : 0 %