KROHNE MFC 010 C Converter User Manual
Page 68
66
MFC010 Interface Manual
Register No.
Description
3085 / 3086
User Defined Volume Total Units Scaling
- The scaling factor used for the user defined Volume Total units ( see
Holding Register No. 1023)
. See Section 8.7 on page 77 for more details.
Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1
3087 / 3088
User Defined Mass Flow Units Scaling
- The scaling factor used for the user defined Mass Flow units ( see Holding
Register No. 1020)
. See Section 8.7 on page 77 for more details.
Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1
3089 / 3090
User Defined Volume Flow Units Scaling
- The s caling f actor us ed f or t he us er defined Volume Flow units (see
Holding Register No. 1024)
. See Section 8.7 on page 77 for more details.
Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1
3091 / 3092
User Defined Density Units Scaling
- The scaling factor used for the user defined Density units ( see Holding
Register No. 1021)
. See Section 8.7 on page 77 for more details.
Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1
3093 / 3094
Calibration Density
- The D ensity v alue us ed by t he c alibration pr ocess w hen “ Other“ i s
selected as the calibration type. T his v alue i s s caled ac cording t o t he
setting of the Density Units (see Holding Register No. 1021). See Section
8.2 on page 72 for more details.
Format : Floating Point
Range : 0 kg/m
3
to 3000 kg/m
3
Default Value : 0 kg/m
3
3095 / 3096
RO
Temperature at Last Zero Calibration
- A stored value of the temperature reading during the last zero Mass flow
calibration. The value is scaled according to the setting of the
“Temperature Units” (see Holding Register No.1025)
Format : Floating Point
Range : -200 °C to + 500 °C