MFC010 Interface Manual

Register No.

Description

3085 / 3086

User Defined Volume Total Units Scaling

- The scaling factor used for the user defined Volume Total units ( see

Holding Register No. 1023)

. See Section 8.7 on page 77 for more details.

Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1

3087 / 3088

User Defined Mass Flow Units Scaling

- The scaling factor used for the user defined Mass Flow units ( see Holding

Register No. 1020)

. See Section 8.7 on page 77 for more details.

Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1

3089 / 3090

User Defined Volume Flow Units Scaling

- The s caling f actor us ed f or t he us er defined Volume Flow units (see

Holding Register No. 1024)

. See Section 8.7 on page 77 for more details.

Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1

3091 / 3092

User Defined Density Units Scaling

- The scaling factor used for the user defined Density units ( see Holding

Register No. 1021)

. See Section 8.7 on page 77 for more details.

Format : Floating Point
Range : 0.00000001 to 1,000
Default Value : 1

3093 / 3094

Calibration Density

- The D ensity v alue us ed by t he c alibration pr ocess w hen “ Other“ i s

selected as the calibration type. T his v alue i s s caled ac cording t o t he

setting of the Density Units (see Holding Register No. 1021). See Section

8.2 on page 72 for more details.

Format : Floating Point
Range : 0 kg/m

3

to 3000 kg/m

3

Default Value : 0 kg/m

3

3095 / 3096

RO

Temperature at Last Zero Calibration

- A stored value of the temperature reading during the last zero Mass flow

calibration. The value is scaled according to the setting of the

“Temperature Units” (see Holding Register No.1025)

Format : Floating Point
Range : -200 °C to + 500 °C