KROHNE MFC 010 C Converter User Manual
Page 49
MFC010 Interface Manual
47
Register No.
Description
3045 / 3046
Density Calibration Coefficient DCF8
- The Temperature Correction value calculated for the density calibration point #2
Format : Floating Point
Range : 0.01 to 10 000 000
3047 / 3048
Zero Calibration Percent
- The Zero Calibration value represented as a percentage of nominal flow
(determined by the sensor type)
Format : Floating Point
Range : -10% to +10%
3049 / 3050
Maximum Recorded Instrument Temperature
- The maximum process temperature that the sensor has measured. The value is
scaled according to the setting of the “Temperature Units” (see Holding Register
No.1025)
Format : Floating Point
Range : +20°C to + 500 °C
3051 / 3052
Minimum Recorded Instrument Temperature
- The minimum process temperature that the sensor has measured. The value is
scaled according to the setting of the “Temperature Units” (see Holding Register
No.1025)
Format : Floating Point
Range : -250 °C to + 20°C
3053 / 3054
2 Phase Signal
- Indication of the 2 phase signal level of the unit. This value is application & process
dependent and it can be used to determine the set point for the 2-phase flow alar,m
function ( see Holding Register No.3021)
Range : 0.0 to 1000.0
5001 - 5004
Mass Total
- The accumulation of mass flow over time. This totaliser can be reset using the
“Reset Totalisers” command (see Coil Register No.1003). The value transmitted is
scaled according to the setting of the “Mass Total Units” (see Holding Register
No.1022)
Format : Double Precision Floating Point
Range : - 1 x 10
12
to +1 x 10
12
kg
5005 - 5008
Volume Total
- The accumulation of volume flow over time. This totaliser can be reset using the
“Reset Totalisers” command (see Coil Register No.1003). The value transmitted is
scaled according to the setting of the “Volume Total Units” (see Holding Register
No.1023)
Format : Double Precision Floating Point
Range : - 1 x 10
9
to +1 x 10
9
m
3
5009 - 5012
Concentration 1 Total
- The accumulation of mass of the concentration product over time. This totaliser
can be reset using the “Reset Totalisers” command (see Coil Register No.1003).
The value transmitted is scaled according to the setting of the “Mass Total Units”
(see Holding Register No.1022) or “Volume Flow Units (see Holding Register 1023).
NB Input register 5009 contains a reading with higher resolution, double floating
point format, for this variable.
Format : Double Precision Floating Point
Range : -1 x 10
12
to +1 x 10
12
kg or -1 x 10
9
to +1 x 10
9
m
3