MFC010 Interface Manual

Bit

Error Flag Designation

17

CT

Name

:

Drive System Open Circuit

Cause

:

The tube oscillation control system has detected an open circuit in

the connection the to tube drive coil.

Action :

Call service support engineer.

18

CT

Name

:

Drive System Short Circuit

Cause

:

The tube oscillation control system has detected an short circuit in

the connection to the tube drive coil.

Action :

Call service support engineer.

19

CT

Name

:

Front End Amplifier Failure

Cause

:

The measurement system has detected a failure in the motion

sensor interface circuit.

Action :

Switch off the unit immediately

to prevent damage to the sensor.

This cannot be repaired in the field, the Front End electronics must be

replaced. Call a service support engineer.

20

Name

:

Harmonic Distortion of Sensor A

Cause

:

Not Currently Implemented

Action :

Call service support engineer

21

Name

:

Harmonic Distortion of Sensor B

Cause

:

Not Currently Implemented

Action :

Call service support engineer

22

CT

Name

:

Watchdog Reset Event

Cause

:

Not Currently Implemented

Action :

Call service support engineer

23

Name

:

Reserved for Future Use

24

Name

:

Reserved for Future Use

25

Name

:

Reserved for Future Use

26

Name

:

Reserved for Future Use

27

Name

:

Reserved for Future Use

28

Name

:

Reserved for Future Use

29

Name

:

Reserved for Future Use

30

Name

:

Reserved for Future Use

31

Name

:

Reserved for Future Use