MFC010 Interface Manual

73

The procedure for performing a single point density calibration is as follows.

An error flag is set in the “Density Calibration” input status register when the measured tube frequency

does not correlate with the value/product type of the calibration density entered by the user ( e.g. if the

user requests a calibration of the “Air” type when the tube is full of process fluid.)

Set the Process

Density

Set the Product type

(Holding Reg. 1028)

Product Type = 3 ?

YES

NO

Set the Density Value

(Holding Reg. 3093)

Start Single Point

Calibration

(Set Output Reg. 1011)

Density Calibration

Status = 1 ?

(Input Reg 1007)

YES

NO

Error During Density

Calibration (See notes)

Store Calibration

(Set Output Register 1001)