7 07 (0716): read exception status, Read exception status – KROHNE MFC 010 C Converter User Manual
Page 31
MFC010 Interface Manual
29
6.7
07 (07
16
): Read Exception Status
When the Master device requests this command function, the MFC010 will respond with a single 8 bit
data character summarizing the status of the instrument. The Master query telegram format is.
Request
Character
Field
For Example
1
Slave Address
01
16
Request to Slave ID 1
2
Function
07
16
“Read Exception Status”
3
CRC Lo
41
16
CRC Checksum
4
CRC Hi
E2
16
The MFC010 will respond to such a request with a telegram formatted as follows.
Response
Character
Field
For Example
1
Slave Address
01
16
Response from Slave ID 1
2
Function
07
16
“Read Exception Status”
3
Status
3D
16
Data = 00111101
2
4
CRC Lo
E3
16
CRC Checksum
5
CRC Hi
E1
16
The Status character received in the response will be formatted as follows :
Bit 0
(LSB)
System State :
00
2
= Measuring,
01
2
= Standby,
10
2
= Stop,
11
2
= Start-up.
Bit 1
Bit 2
EEPROM Save Status :
0 = All Data Saved to EEPROM
1= Data Write to EEPOM Pending
Bit 3
Process Control Status :
0 = Process Control Inactive (Control Condition Invalid)
1 = Process Control Active (Control Condition Valid)
Bit 4
Zero Calibration Status :
0 = Zero Calibration OK
1 = Zero Calibration Error
Bit 5
Density Calibration Status : 0 = Density Calibration OK
1 = Density Calibration Error
Bit 6
Process Warning Status : 0 = No Process Warning Flag(s) Detected
1 = Process Warning Flag(s) Detected
Bit 7 (MSB) System Error Status :
0 = No System Error Flag(s) Detected
1 = System Error Condition Flag(s) Detected