MFC010 Interface Manual

53

Register No.

Description

1015

Tube Amplitude

- The sensor input amplitude to which the measuring tube oscillation control

system must aim to achieve. T he optimum setting for this value depends

upon the sensor type and size.

Format : Unsigned Integer
Range : 1% to 98%
Default Value: Dependent on Sensor Type and Size.

1016

Concentration Type

- The extrapolation method used by the General concentration

measurement c alculation s ub-system. (See Section 8.4 on page 76 for

more details)

Format : Unsigned Integer
Range :

0 = Linear

1 = Non-Linear

Default Value : 0

1017

Concentration 1 Product

- Determines the method by which the concentration value is represented

when a mixture two products, A & B, are being measured. (See Section 8.4

on page 76 for more details)

Format : Unsigned Integer
Range :

0 = % of Product A

1 = % of Product B

Default Value : 0

1018

Concentration Coefficient #5

- Definition of the carrier ( Product B ) w hen per forming t he G eneral

Concentration calculation. (See section 8.4 on page 76 for details)

Format : Unsigned Integer
Range:

0 = Pure Water (998.2 kg/m

3

@ 20°C)

1 = Town Water (999.7 kg/m

3

@ 20°C)

2 = Other.

Default Value : 0

1019

CT

Density Mode

- The mode by which the density value is generated. S ee section 8.3 on

page 75 for details.

Format : Unsigned Integer
Range:

0 = Actual (Measured Density)

1 = Fixed (Entered at register No. 3057)

2 = Referred.

Default Value : 0