MFC010 Interface Manual
45
Register No.
Description
3015 / 3016
Concentration 2
- The proportion of the selected product in the solution. The value transmitted is
scaled according to the setting of the “Concentration 2 Function” (see Holding
Register No.1029). When using “% Mass” or “% Volume” the selected product may
be either the solute or the solvent according to the setting of “Concentration 2
Product” (see Holding Register No.1030).
Format : Floating Point
Range : 0.0% to 100.0%
3017 / 3018
Velocity
- The velocity at which the medium is passing through the sensor, determined by the
Volume Flow and the Pipe Diameter (see Holding Register No.3043). The value
transmitted is scaled according to the setting of the “Velocity Units” (see Holding
Register No.1026)
Format : Floating Point
Range : Dependant on the sensor type and selected units
3019 / 3020
Mass Total
- The accumulation of mass flow over time. This totaliser can be reset using the
“Reset Totalisers” command (see Coil Register No.1003). The value transmitted is
scaled according to the setting of the “Mass Total Units” (see Holding Register
No.1022). NB Input register 5001 contains a reading with higher resolution, double
floating point format, for this variable.
Format : Floating Point
Range : -1 x 10
12
to +1 x 10
12
kg
3021 / 3022
Volume Total
- The accumulation of volume flow over time. This totaliser can be reset using the
“Reset Totalisers” command (see Coil Register No.1003). The value transmitted is
scaled according to the setting of the “Volume Total Units” (see Holding Register
No.1023) ). NB Input register 5005 contains a reading with higher resolution, double
floating point format, for this variable.
Format : Floating Point
Range : -1 x 10
9
to +1 x 10
9
m
3
3023/ 3024
Concentration 1 Total
- The accumulation of mass of the concentration product over time. This totaliser
can be reset using the “Reset Totalisers” command (see Coil Register No.1003).
The value transmitted is scaled according to the setting of the “Mass Total Units”
(see Holding Register No.1022) or “Volume Flow Units (see Holding Register 1023).
NB Input register 5009 contains a reading with higher resolution, double floating
point format, for this variable.
Format : Floating Point
Range : -1 x 10
12
to +1 x 10
12
kg or -1 x 10
9
to +1 x 10
9
m
3
3025 / 3026
Concentration 2 Total
- The accumulation of mass of the concentration product over time. This totaliser
can be reset using the “Reset Totalisers” command (see Coil Register No.1003).
The value transmitted is scaled according to the setting of the “Mass Total Units”
(see Holding Register No.1022) or “Volume Flow Units (see Holding Register 1023).
NB Input register 5013 contains a reading with higher resolution, double floating
point format, for this variable.
Format : Floating Point
Range : -1 x 10
12
to +1 x 10
12
kg or -1 x 10
9
to +1 x 10
9
m
3