12

MFC010 Interface Manual


NB

For H azardous A rea appl ications t he us er s hould r efer t o s ection 4.3, on page 17, f or c onnection

details.


NB

For multidrop systems, ensure cycle times are properly calculated to ensure bus speeds are adequate

for the application.

3.5

Installation Guidelines for Electromagnetic Compatibility

Whilst t he M FC010 and i ts as sociated s ensor has been des igned, t ested and c ertified t o m eet the

requirements on international s tandards of E lectromagnetic C ompatibility ( EMC), i t i s t he us ers

responsibility to ensure that the connection guidelines described in this document are followed. In addition

the user should use recognised good practise in the location and cable routing of the MFC010 in relation

to its surrounding environment. The user should consider the following suggestions when installing an

MFC010 into a system.

1. Every effort should be made to avoid significant lengths (>50mm) of unshielded signal wire

when connecting to the system, any terminal connections should be housed in a suitably

shielded enclosure.

2. Avoid routing the cable in groups with or alongside other power carrying cables.

3. Avoid locating the MFC010 or routing the connection cable in close proximity to large

electrically powered equipment, such as pumps, inverters etc.

4. If necessary, route the connection cable through a suitably earthed metal conduit.