MFC010 Interface Manual

51

Register No.

Description

1010

CT

Process Control Condition

- The value stored in this register determines the source measurement that

is used to activate the Internal Process Control System. See section 8.8 on

page 78 for further details.

Format : Unsigned Integer
Range :

0 = Density

1 = Temperature

Default Value : 1

1011

Concentration 1 Function

- Determines the type of measurement calculation that the meter performs for

Concentration 1. Can only be activated by means of the Concentration Password

register No.7007. See section 8.4 on page 76 for details.

Format : Unsigned Integer

Range :

0 = De-activated (Option NOT Installed)

1 = Off

2 = °Brix

3 = % Mass (General Concentration)

4 = °Baumé 144.3

5 = °Baumé 145.0

6 = % NaOH

7 = °Plato

8 = % Volume (General Concentration)

9 = °API
Default Value : 0

1012

CT

Sensor Type

- The setting of the Sensor type determines the available sizes, materials

and oper ating par ameters of t he s ensor t o which t he MFC10 is attached.

Changing this value will def ault t he S ensor S ize ( Register N o. 1013) and

depending on option chosen the Material (Register No. 1014). WARNING :

Changing this value will reset the calibration coefficients.

Format : Unsigned Integer
Range :

0 = OPTIMASS 7000

1 = OPTIMASS 3000

2 = OPTIGAS 5000

3 = OPTIMASS 8000

4 = OPTIMASS 9000

5 = OPTIMASS 1000

6 = OPTIMASS 2000

9 = OPTIGAS 4000


Default Value : Dependent on meter supplied