KROHNE MFC 010 C Converter User Manual
Page 53
MFC010 Interface Manual
51
Register No.
Description
1010
CT
Process Control Condition
- The value stored in this register determines the source measurement that
is used to activate the Internal Process Control System. See section 8.8 on
page 78 for further details.
Format : Unsigned Integer
Range :
0 = Density
1 = Temperature
Default Value : 1
1011
Concentration 1 Function
- Determines the type of measurement calculation that the meter performs for
Concentration 1. Can only be activated by means of the Concentration Password
register No.7007. See section 8.4 on page 76 for details.
Format : Unsigned Integer
Range :
0 = De-activated (Option NOT Installed)
1 = Off
2 = °Brix
3 = % Mass (General Concentration)
4 = °Baumé 144.3
5 = °Baumé 145.0
6 = % NaOH
7 = °Plato
8 = % Volume (General Concentration)
9 = °API
Default Value : 0
1012
CT
Sensor Type
- The setting of the Sensor type determines the available sizes, materials
and oper ating par ameters of t he s ensor t o which t he MFC10 is attached.
Changing this value will def ault t he S ensor S ize ( Register N o. 1013) and
depending on option chosen the Material (Register No. 1014). WARNING :
Changing this value will reset the calibration coefficients.
Format : Unsigned Integer
Range :
0 = OPTIMASS 7000
1 = OPTIMASS 3000
2 = OPTIGAS 5000
3 = OPTIMASS 8000
4 = OPTIMASS 9000
5 = OPTIMASS 1000
6 = OPTIMASS 2000
9 = OPTIGAS 4000
Default Value : Dependent on meter supplied