MFC010 Interface Manual

Register No.

Description

3001 / 3002

Mass Flow
- The measured mass flow rate after filtering. Proportional to the Phase shift

detected between the sensors. The value transmitted is scaled according to the

setting of the “Mass Flow Units” (see Holding Register No.1020)
Format : Floating Point
Range : Dependant on the sensor type and selected units

3003 / 3004

Density
- The measured density after filtering. Inversely proportional to the oscillation

frequency of the measuring tube. The value transmitted is scaled according to the

setting of the “Density Units” (see Holding Register No.1021). This value may be

fixed or referred according to the setting of the ”Density Mode” (see Holding Register

No.1019)
Format : Floating Point
Range : 0.05 kg/m³ to 3000 kg/m³

3005 / 3006

Temperature
- The measured temperature of the measuring tube. The value is scaled according

to the setting of the “Temperature Units” (see Holding Register No.1025)
Format : Floating Point
Range : -250 °C to + 500 °C

3007 / 3008

Volume Flow
- The volume flow rate, determined from the Mass flow and Density measurements.

The value transmitted is scaled according to the setting of the “Volume Flow Units”

(see Holding Register No.1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units

3009 / 3010

Concentration 1 Flow
- The flowrate of the defined concentration component. The value transmitted is

scaled according to the setting of the “Mass Flow Units” (see Holding Register

No.1020) or “Volume Flow Units ( see Holding Register No. 1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units

3011 / 3012

Concentration 2 Flow
- The flowrate of the defined concentration component. The value transmitted is

scaled according to the setting of the “Mass Flow Units” (see Holding Register

No.1020) or “Volume Flow Units ( see Holding Register No. 1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units

3013 / 3014

Concentration 1
- The proportion of the selected product in the solution. The value transmitted is

scaled according to the setting of the “Concentration 1 Function” (see Holding

Register No.1011). When using “% Mass” or “% Volume” the selected product may

be either the solute or the solvent according to the setting of “Concentration 1

Product” (see Holding Register No.1017).
Format : Floating Point
Range : Dependant on the selected units