KROHNE MFC 010 C Converter User Manual
Page 46
44
MFC010 Interface Manual
Register No.
Description
3001 / 3002
Mass Flow
- The measured mass flow rate after filtering. Proportional to the Phase shift
detected between the sensors. The value transmitted is scaled according to the
setting of the “Mass Flow Units” (see Holding Register No.1020)
Format : Floating Point
Range : Dependant on the sensor type and selected units
3003 / 3004
Density
- The measured density after filtering. Inversely proportional to the oscillation
frequency of the measuring tube. The value transmitted is scaled according to the
setting of the “Density Units” (see Holding Register No.1021). This value may be
fixed or referred according to the setting of the ”Density Mode” (see Holding Register
No.1019)
Format : Floating Point
Range : 0.05 kg/m³ to 3000 kg/m³
3005 / 3006
Temperature
- The measured temperature of the measuring tube. The value is scaled according
to the setting of the “Temperature Units” (see Holding Register No.1025)
Format : Floating Point
Range : -250 °C to + 500 °C
3007 / 3008
Volume Flow
- The volume flow rate, determined from the Mass flow and Density measurements.
The value transmitted is scaled according to the setting of the “Volume Flow Units”
(see Holding Register No.1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units
3009 / 3010
Concentration 1 Flow
- The flowrate of the defined concentration component. The value transmitted is
scaled according to the setting of the “Mass Flow Units” (see Holding Register
No.1020) or “Volume Flow Units ( see Holding Register No. 1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units
3011 / 3012
Concentration 2 Flow
- The flowrate of the defined concentration component. The value transmitted is
scaled according to the setting of the “Mass Flow Units” (see Holding Register
No.1020) or “Volume Flow Units ( see Holding Register No. 1024)
Format : Floating Point
Range : Dependant on the sensor type and selected units
3013 / 3014
Concentration 1
- The proportion of the selected product in the solution. The value transmitted is
scaled according to the setting of the “Concentration 1 Function” (see Holding
Register No.1011). When using “% Mass” or “% Volume” the selected product may
be either the solute or the solvent according to the setting of “Concentration 1
Product” (see Holding Register No.1017).
Format : Floating Point
Range : Dependant on the selected units