MFC010 Interface Manual
63
Register No.
Description
3051 / 3052
Process Control Maximum Limit
- This v alue det ermines t he l evel abov e w hich t he c ontrol function will
become active. The scaling and range of this value is dependant upon the
setting of t he c ontrol condition (see Holding R egister N o 1010 ). S ee
Section 8.8 on page 78 for more details.
Format : Floating Point
Range : Dependant on the Control Condition and Scaling Units
Default Value : 2000 kg/m
3
or +100.0°C
3053 / 3054
Process Control Minimum Limit
- This value determines the level below which the control function will
become active. The scaling and range of this value is dependant upon the
setting of t he c ontrol c ondition ( see H olding R egister N o 1010 ). S ee
Section 8.8 on page 78 for more details.
Format : Floating Point
Range : Dependant on the Control Condition and Scaling Units
Default Value : 500 kg/m
3
or 0.0°C
3055 / 3056
Referred Density Reference Temperature
- The reference temperature to which density reading is referred when the
Density Mode (see Holding Register 1019) is set to “Referred”. This value
is scaled by the setting of the Temperature Units (see holding register No.
1025)
. Refer to Section 8.3 on page 75 for further details on the use of the
Referred Density Mode.
Format : Floating Point
Range : -200 °C to + 500 °C
Default Value : +20 °C
3057/ 3058
Fixed Density Value
- The fixed value of the density that i s us ed by t he c ompensation and
concentration calculations when the D ensity M ode ( see H olding R egister
1019) is set to “Fixed”. T his value is scaled by the setting of the Density
Units (see holding register No. 1021). Refer to Section 8.3 on page 75 for
further details on the use of the Fixed Density Mode.
Format : Floating Point
Range : 0.08 kg/m
3
to 3000 kg/m
3
Default Value : 998.2 kg/m
3
3059 / 3060
Referred Density Slope
- The slope used to extrapolate the density from the measured temperature
to the referred temperature when the Density Mode (see Holding Register
1019)
is set to “Referred”. This value is scaled by the setting of the Density
Units (see holding register No. 1021) and the Temperature Units (see
holding r egister N o. 1025) . R efer t o S ection 8.3 on page 75 for f urther
details on the use of the Referred Density Mode.
Format : Floating Point
Range : 0 kg/m
3
/°C to 65 kg/m
3
/°C
Default Value : 0 kg/m
3
/°C