MFC010 Interface Manual

63

Register No.

Description

3051 / 3052

Process Control Maximum Limit

- This v alue det ermines t he l evel abov e w hich t he c ontrol function will

become active. The scaling and range of this value is dependant upon the

setting of t he c ontrol condition (see Holding R egister N o 1010 ). S ee

Section 8.8 on page 78 for more details.

Format : Floating Point
Range : Dependant on the Control Condition and Scaling Units
Default Value : 2000 kg/m

3

or +100.0°C

3053 / 3054

Process Control Minimum Limit

- This value determines the level below which the control function will

become active. The scaling and range of this value is dependant upon the

setting of t he c ontrol c ondition ( see H olding R egister N o 1010 ). S ee

Section 8.8 on page 78 for more details.

Format : Floating Point
Range : Dependant on the Control Condition and Scaling Units
Default Value : 500 kg/m

3

or 0.0°C

3055 / 3056

Referred Density Reference Temperature

- The reference temperature to which density reading is referred when the

Density Mode (see Holding Register 1019) is set to “Referred”. This value

is scaled by the setting of the Temperature Units (see holding register No.

1025)

. Refer to Section 8.3 on page 75 for further details on the use of the

Referred Density Mode.

Format : Floating Point
Range : -200 °C to + 500 °C
Default Value : +20 °C

3057/ 3058

Fixed Density Value

- The fixed value of the density that i s us ed by t he c ompensation and

concentration calculations when the D ensity M ode ( see H olding R egister

1019) is set to “Fixed”. T his value is scaled by the setting of the Density

Units (see holding register No. 1021). Refer to Section 8.3 on page 75 for

further details on the use of the Fixed Density Mode.

Format : Floating Point
Range : 0.08 kg/m

3

to 3000 kg/m

3

Default Value : 998.2 kg/m

3

3059 / 3060

Referred Density Slope

- The slope used to extrapolate the density from the measured temperature

to the referred temperature when the Density Mode (see Holding Register

1019)

is set to “Referred”. This value is scaled by the setting of the Density

Units (see holding register No. 1021) and the Temperature Units (see

holding r egister N o. 1025) . R efer t o S ection 8.3 on page 75 for f urther

details on the use of the Referred Density Mode.

Format : Floating Point
Range : 0 kg/m

3

/°C to 65 kg/m

3

/°C

Default Value : 0 kg/m

3

/°C