MFC010 Interface Manual

83

Bit

Error Flag Designation

9

CT

Name

:

MT Strain Gauge Open Circuit

Cause

:

The measurement system has detected an electrical open circuit

between its measuring tube strain gauge input terminals,

therefore strain measurement is not possible. The system will

default to a static strain reading of 1000Ω when this error occurs,

however, internally it will use the calibration reference strain for

measurement calculation to prevent sudden massive variation in

the mass flow or density measurements.

Action :

Call a service support engineer

10

CT

Name

:

IC Strain Gauge Short Circuit

Cause

:

The measurement system has detected an electrical short circuit

between its inner cylinder strain gauge input terminals, therefore

strain measurement is not possible. The system will default to a

static strain reading of 0Ω when this error occurs, however,

internally it will use the calibration reference strain for

measurement calculation to prevent sudden massive variation in

the mass flow or density measurements.

Action :

Call service support engineer

11

CT

Name

:

IC Strain Gauge Open Circuit

Cause

:

The measurement system has detected an electrical open circuit

between its inner cylinder strain gauge input terminals, therefore

strain measurement is not possible. The system will default to a

static strain reading of 1000Ω when this error occurs, however,

internally it will use the calibration reference strain for

measurement calculation to prevent sudden massive variation in

the mass flow or density measurements.

Action :

Call service support engineer

12

CT

Name

:

Sensor A

Cause

:

The measured level of Sensor A is below 5% of the required value.

This error can occur for short periods following rapid changes of product

density, such as when the tube changes to or from an empty state or

when excessive Air bubbles are present in the product.

Action :

If this Error is persistently reported during normal operation call your

service support engineer for further advice.

13

CT

Name

:

Sensor B

Cause

:

The measured level of Sensor B is below 5% of the required value.

This error can occur for short periods following rapid changes of product

density, such as when the tube changes to or from an empty state or

when excessive Air bubbles are present in the product.

Action :

If this Error is persistently reported during normal operation call your

service support engineer for further advice.

14

CT

Name

:

Ratio A/B

Cause

:

The ratio of the two sensor signals is too large, i.e. one of the

motion sensors is giving a reading significantly below the other

sufficient to cause an error in the mass flow calculations. This

may be caused by or the cause of instability in the tube oscillation

control system.

Action :

Call service support engineer

15

CT

Name

:

DC A

Cause

:

The DC component of the motion Sensor A signal is too large.

Action :

Call service support engineer.

16

CT

Name

:

DC B

Cause

:

The DC component of the motion Sensor B signal is too large.

Action :

Call service support engineer.