MFC010 Interface Manual
Register No.
Description
1027
Additional Totaliser Source
- The additional totaliser acts as a bat ch c ounter r ecording a s ub-total o f
one of the four main totalisers. The setting of this register determines which
of t he f our m ain t otalisers i s t he s ource for the Additional T otaliser. T he
Additional Totaliser can be r eset individually from the main totalisers using
Output Status Register 1004. The Additional Totaliser is also reset when
the main totalisers are reset using Output Status register 1003.
Format : Unsigned Integer
Range :
0 = Disabled
1 = Mass Total
2 = Volume Total
3 = Concentration 1 Total
4 = Concentration 2 Total
Default Value : 0
1028
Density Calibration Product Type
- The fluid type defined for the Density calibration commands (See Output
Status Register Nos. 1011-1013). S ee Section 8.2 on page 72 for further
details about the density calibration procedure.
Format : Unsigned Integer
Range :
0 = Empty
1 = Pure Water (998.2 kg/m
3
@ 20°C )
2 = Town Water (999.7 kg/m
3
@ 20°C )
3 = Other
Default Value : 0
1029
Concentration 2 Function
- Determines the type of measurement calculation that the meter performs for
Concentration 2. Can only be activated by means of the Concentration Password
register No.7007. See section 8.4 on page 76 for details.
Format : Unsigned Integer
Range :
0 = De-activated (Option NOT Installed)
1 = Off
2 = °Brix
3 = % Mass (General Concentration)
4 = °Baumé 144.3
5 = °Baumé 145.0
6 = % NaOH
7 = °Plato
8 = % Volume (General Concentration)
9 = °API
Default Value : 0
1030
Concentration 2 Product
- Determines the method by which the concentration value is represented
when a mixture two products, A & B, are being measured. (See Section 8.4
on page 76 for more details)
Format : Unsigned Integer
Range :
0 = % of Product A
1 = % of Product B
Default Value : 0