beautypg.com

KROHNE MFC 010 C Converter User Manual

Page 60

background image


58

MFC010 Interface Manual

Register No.

Description

1027

Additional Totaliser Source

- The additional totaliser acts as a bat ch c ounter r ecording a s ub-total o f

one of the four main totalisers. The setting of this register determines which

of t he f our m ain t otalisers i s t he s ource for the Additional T otaliser. T he

Additional Totaliser can be r eset individually from the main totalisers using

Output Status Register 1004. The Additional Totaliser is also reset when

the main totalisers are reset using Output Status register 1003.

Format : Unsigned Integer
Range :

0 = Disabled

1 = Mass Total

2 = Volume Total

3 = Concentration 1 Total

4 = Concentration 2 Total

Default Value : 0

1028

Density Calibration Product Type

- The fluid type defined for the Density calibration commands (See Output

Status Register Nos. 1011-1013). S ee Section 8.2 on page 72 for further

details about the density calibration procedure.

Format : Unsigned Integer
Range :

0 = Empty

1 = Pure Water (998.2 kg/m

3

@ 20°C )

2 = Town Water (999.7 kg/m

3

@ 20°C )

3 = Other

Default Value : 0

1029

Concentration 2 Function

- Determines the type of measurement calculation that the meter performs for

Concentration 2. Can only be activated by means of the Concentration Password

register No.7007. See section 8.4 on page 76 for details.

Format : Unsigned Integer

Range :

0 = De-activated (Option NOT Installed)

1 = Off

2 = °Brix

3 = % Mass (General Concentration)

4 = °Baumé 144.3

5 = °Baumé 145.0

6 = % NaOH

7 = °Plato

8 = % Volume (General Concentration)

9 = °API
Default Value : 0

1030

Concentration 2 Product

- Determines the method by which the concentration value is represented

when a mixture two products, A & B, are being measured. (See Section 8.4

on page 76 for more details)

Format : Unsigned Integer
Range :

0 = % of Product A

1 = % of Product B

Default Value : 0