MFC010 Interface Manual

43

7.4

Input Registers

The input registers contain the measurement values and di agnostic data that the MFC010 produces. All

of these registers are read only.

Register No.

Description

1001

Sensor A Level
- The measured input level for Sensor A as a percentage of the maximum possible

input.
Format : Unsigned Integer
Range : 0% to 100%

1002

Sensor B Level
– The measured input level for Sensor B as a percentage of the maximum possible

input.
Format : Unsigned Integer
Range : 0% to 100%

1003

Drive Level
– The output power to the Drive coil as a percentage of the maximum possible

output.
Format : Unsigned Integer
Range : 0% to 100%

1004

System State
- The current operating condition of the sensor system. This can be changed using

the “Request System State” commands (see Coil Register No’s 1005 – 1007).
Format : Unsigned Integer
Range :

1 = Stop

2 = Start-up

3 = Measuring

4 = Not Used

5 = Standby

6 = Zero Calibration

1005

Density Calibration Coefficient DCF1
- The fluid type defined for the calibration of Density Point #1
Format : Unsigned Integer
Range :

0 = Empty

1 = Pure Water (998.2 kg/m

3

@ 20°C )

2 = Town Water (999.7 kg/m

3

@ 20°C )

3 = Other

1006

Density Calibration Coefficient DCF2
- The fluid type defined for the calibration of Density Point #2
Format : Unsigned Integer
Range :

0 = Empty

1 = Pure Water (998.2 kg/m

3

@ 20°C )

2 = Town Water (997.7 kg/m

3

@ 20°C )

3 = Other