4 input registers, Input registers – KROHNE MFC 010 C Converter User Manual
Page 45

MFC010 Interface Manual
43
7.4
Input Registers
The input registers contain the measurement values and di agnostic data that the MFC010 produces. All
of these registers are read only.
Register No.
Description
1001
Sensor A Level
- The measured input level for Sensor A as a percentage of the maximum possible
input.
Format : Unsigned Integer
Range : 0% to 100%
1002
Sensor B Level
– The measured input level for Sensor B as a percentage of the maximum possible
input.
Format : Unsigned Integer
Range : 0% to 100%
1003
Drive Level
– The output power to the Drive coil as a percentage of the maximum possible
output.
Format : Unsigned Integer
Range : 0% to 100%
1004
System State
- The current operating condition of the sensor system. This can be changed using
the “Request System State” commands (see Coil Register No’s 1005 – 1007).
Format : Unsigned Integer
Range :
1 = Stop
2 = Start-up
3 = Measuring
4 = Not Used
5 = Standby
6 = Zero Calibration
1005
Density Calibration Coefficient DCF1
- The fluid type defined for the calibration of Density Point #1
Format : Unsigned Integer
Range :
0 = Empty
1 = Pure Water (998.2 kg/m
3
@ 20°C )
2 = Town Water (999.7 kg/m
3
@ 20°C )
3 = Other
1006
Density Calibration Coefficient DCF2
- The fluid type defined for the calibration of Density Point #2
Format : Unsigned Integer
Range :
0 = Empty
1 = Pure Water (998.2 kg/m
3
@ 20°C )
2 = Town Water (997.7 kg/m
3
@ 20°C )
3 = Other