MFC010 Interface Manual

Register No.

Description

3027 / 3028

Additional Total
- The additional totaliser can be selected to be a sub-totaliser of one of the other four

main totalisers, allowing batch operation and reset without affecting the main totaliser

values. Use Holding register No 1027 to select which of the four main totalisers is

the source totaliser. This totaliser can be reset using the “Reset Totalisers”

command (see Coil Register No.1003) or the “Reset Additional Totaliser” command

(see Coil Register No 1004). The value transmitted is scaled according to the setting

of the units associated with the source totaliser. NB Input register 5017 contains a

reading with higher resolution, double floating point format, for this variable.
Format : Floating Point
Range : -1 x 10

12

to +1 x 10

12

kg or -1 x 10

9

to +1 x 10

9

m

3

3029 / 3030

Tube Frequency
- The measured oscillation frequency of the measuring tube
Format : Floating Point
Range : 50Hz to 2000 Hz

3031 / 3032

Measuring Tube Strain
- The measured resistance of the measuring tube strain gauge.
Format : Floating Point
Range : 0 Ω to 1000 Ω

3033 / 3034

Inner Cylinder Strain
- The measured resistance of the inner cylinder strain gauge.
Format : Floating Point
Range : 0 Ω to 1000 Ω

3035 / 3036

Density Calibration Coefficient DCF2
- The density value for the density calibration point #1
Format : Floating Point
Range : -1000 kg/m³ to 3000 kg/m³

3037 / 3038

Density Calibration Coefficient DCF3
- The Density Flow Compensation value calculated for the density calibration point

#1
Format : Floating Point
Range : 0.01 to 100

3039 / 3040

Density Calibration Coefficient DCF4
- The Temperature Correction value calculated for the density calibration point #1
Format : Floating Point
Range : 0.01 to 10 000 000

3041 / 3042

Density Calibration Coefficient DCF6
- The density value for the density calibration point #2
Format : Floating Point
Range : -1000 kg/m³ to 3000 kg/m³

3043 / 3044

Density Calibration Coefficient DCF7
- The Density Flow Compensation value calculated for the density calibration point

#2
Format : Floating Point
Range : 0.01 to 100