KROHNE MFC 010 C Converter User Manual

Page 44

background image


42

MFC010 Interface Manual

Register No.

Description

1007

Density Calibration Status

– This state indicates when the system has detected an error during the density

calibration. This will only be active for the latest attempt at a density calibration. If

this bit indicates that the latest density calibration failed, then the previous density

calibration will still be being used. See Section 8.2 on page 72 for further details on

Density calibration.

Range :

0 = The last Density Calibration was successful

1 = An error was detected during the last Density Calibration

1008

Mass Flow Zero Calibration Status

– This state indicates when the system has detected an error during the mass flow

zero calibration. This will only be active for the latest attempt at a calibration. If this

bit indicates that the latest zero calibration failed, then the previous zero calibration

will still be being used. See Section 8.1 on page 71 for further details on Mass Flow

Zero Calibration.

Range :

0 = The last Mass Flow Zero Calibration was successful

1 = An error was detected during the last Zero Calibration.

1009

Process Control Status

– This state indicates when the Internal process control mechanism is active. See

section 8.8 on page 78 for more details on the Internal Process control system

Range :

0 = Internal Process Control In-active

1 = Internal Process Control Active