KROHNE MFC 010 C Converter User Manual
Page 44
42
MFC010 Interface Manual
Register No.
Description
1007
Density Calibration Status
– This state indicates when the system has detected an error during the density
calibration. This will only be active for the latest attempt at a density calibration. If
this bit indicates that the latest density calibration failed, then the previous density
calibration will still be being used. See Section 8.2 on page 72 for further details on
Density calibration.
Range :
0 = The last Density Calibration was successful
1 = An error was detected during the last Density Calibration
1008
Mass Flow Zero Calibration Status
– This state indicates when the system has detected an error during the mass flow
zero calibration. This will only be active for the latest attempt at a calibration. If this
bit indicates that the latest zero calibration failed, then the previous zero calibration
will still be being used. See Section 8.1 on page 71 for further details on Mass Flow
Zero Calibration.
Range :
0 = The last Mass Flow Zero Calibration was successful
1 = An error was detected during the last Zero Calibration.
1009
Process Control Status
– This state indicates when the Internal process control mechanism is active. See
section 8.8 on page 78 for more details on the Internal Process control system
Range :
0 = Internal Process Control In-active
1 = Internal Process Control Active