60

MFC010 Interface Manual

Register No.

Description

3011 / 3012

CT

Calibration Coefficient CF6

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -10,000,000 to +10,000,000
Default Value : 0

3013 / 3014

CT

Calibration Coefficient CF7

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -10,000,000 to +10,000,000
Default Value : 0

3015 / 3016

CT

Calibration Coefficient CF8

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -10,000,000 to +10,000,000
Default Value : 0

3017 / 3018

RO

Calibration Coefficient CF9

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -10,000 to +10,000
Default Value : 0

3019 / 3020

RO

Calibration Coefficient CF10

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : 1 to 100,000
Default Value : 1000

3021 / 3022

CT

Calibration Coefficient CF11

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -10,000 to +10,000
Default Value : 0

3023 / 3024

CT

Calibration Coefficient CF12

- One of the coefficients used to accurately calculate the values measured

by the sensor, defined during the manufacturing process.

Format : Floating Point
Range : -1,000,000 to +1,000,000
Default Value : 0