KROHNE MFC 010 C Converter User Manual
Page 61
MFC010 Interface Manual
59
Register No.
Description
1031
CT
Calibration Coefficient CF25
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Integer
Range: 0 to 1
Default value: 0
1032
Year of Manufacture
- Year that the mass flowmeter was manufactured. A two digit format is used, starting
at the year 2000
Format: Integer
Range: 0 to 99
Default value: 0
3001 / 3002
CT
Calibration Coefficient CF1
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Floating Point
Range : -40 to +200
Default Value : 20
3003 / 3004
CT
Calibration Coefficient CF2
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Floating Point
Range : 100 to 1000
Default Value : 750
3005 / 3006
CT
Calibration Coefficient CF3
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Floating Point
Range : 100 to 1000
Default Value : 300
3007 / 3008
CT
Calibration Coefficient CF4
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Floating Point
Range : -100,000 to + 100,000
Default Value : 0
3009 / 3010
CT
Calibration Coefficient CF5
- One of the coefficients used to accurately calculate the values measured
by the sensor, defined during the manufacturing process.
Format : Floating Point
Range : 0.0001 to 90,000,000
Default Value : Dependant on Sensor Settings (See Appendix D)