Holding registers – KROHNE MFC 010 C Converter User Manual

Page 37

background image

MFC010 Interface Manual

35

Register

Description

3005

Temperature

3007

Volume Flow

3009

Concentration 1 Flow

3011

Concentration 2 Flow

3013

Concentration 1

3015

Concentration 2

3017

Velocity

3019

Mass Total

3021

Volume Total

3023

Concentration 1 Total

3025

Concentration 2 Total

3027

Additional Total

3029

Tube Frequency

3031

Measuring Tube Strain

3033

Inner Cylinder Strain

3035

DCF2

3037

DCF3

3039

DCF4

3041

DCF6

3043

DCF7

3045

DCF8

3047

Zero Calibration Percent

3049

Maximum Instrument Temperature

3051

Minimum Instrument Temperature

3053

2 Phase Signal

Double Precision Floating Point Values (Accessed in groups of four registers)

5001

Mass Total

5005

Volume Total

5009

Concentration 1 Total

5013

Concentration 2 Total

5017

Additional Total

Long Integer Values (Accessed in pairs of registers)

7001

System Error Flags

7003

Process Warning Flags

7005

Stored System Error Flags

7007

Stored Process Warning Flags

Holding Registers

4XXXX Registers, Accessed Using Commands 03, 06 and 16 (* Command 03 and 06 Only)

Single Registers (16 bit Integer Values)

1001

Supervisor Lock Password*

1002

Service Lock Password*

1003

Custody Transfer Lock Password*

1004

Modbus Communications Format