9.2

Process Warnings

The Process Warning flags, held as a long integer value in input registers 7003/7004, are an indication of

process conditions that are present at that moment that may adversely affect the measurement operation

or accuracy of the mass flow sensor, such as exceeding the specified operating limits of the sensor. Input

registers 7007/7008 hold a stored history of which these warning flags hav e been s et s ince t he l ast

“Reset” command, which can be used to monitor and record the occurrence of transient process warning

conditions. (See Output Coil Number 1009 on page 39 for details of how to reset the stored Error Flags).

Bit

Warning Flag Designation

0

CT

Name

:

Mass Flow Over-range

Cause

:

The measured mass flow is beyond the maximum allowable limit

for this sensor

1

CT

Name

:

Process Temperature Outside Limits

Cause

:

The measured tube temperature has been outside of the

permissible operating limits of the sensor.

2

CT

Name

:

Process Density Over-range

Cause

:

The product density is beyond the allowable limits of the system

to enable accurate measurement

3

CT

Name

:

Mass Totaliser Over-flow

Cause

:

The Mass total value has overflowed, i.e. the mass total variable

has previously reached its maximum permissible value and

therefore has been forced to return to zero after the following

increment.

4

CT

Name

:

Volume Totaliser Over-flow

Cause

:

The Volume total value has overflowed, i.e. the Volume total

variable has previously reached its maximum permissible value

and therefore has been forced to return to zero after the following

increment.

5

Name

:

Additional Totaliser Over-flow

Cause

:

The Additional total value has overflowed, i.e. the Additional total

variable has previously reached its maximum permissible value

and therefore has been forced to return to zero after the following

increment.

6

Name

:

Concentration By Mass Totaliser Over-flow

Cause

:

The Concentration Total - By Mass value has overflowed, i.e. the

Concentration Mass total variable has previously reached its

maximum permissible value and therefore has been forced to

return to zero after the following increment.

7

Name

:

Concentration By Volume Totaliser Over-flow

Cause

:

The Concentration Total – By Volume value has overflowed, i.e.

the Concentration Volume total variable has previously reached

its maximum permissible value and therefore has been forced to

return to zero after the following increment.